Skip to content

5. Assembling mechanical parts, lenses and filters

tmh edited this page Apr 9, 2021 · 5 revisions

Hardware - Firmware - Software

Parts list

❗️All filters must be mounted with the filter layer towards the light source.

⚠️ Do not touch the filter layer!

  • Glue the lens (Edmund 45429 PCX 4x10) to the emission filter with Norland NOA81 optical glue (https://www.norlandprod.com/adhesives/NOA%2081.html) using a glue jig. The lens must be glued on the filter-side because of the direction of the light. Use a UV-light to cure the glue.

Remnants of the glue should be carefully removed with a scalpel.

  • Coat all filters black on the sides with a thin layer of chalkboard paint or a good black permanent marker.

  • Mount the dichroic mirror and excitation filter in the housing.

  • Mount the half-ball lens and slide the sleeve a little bit over the housing.

  • Hold the LED in place and slide the sleeve further over the housing.

  • Mount the emission filter in the housing. It will be in place when the top of the filter is about 2.8mm below the top of the housing.

  • Glue the filter in place with a quick set epoxy glue applied with very small dots at the corners.

  • Pull off the protective foil and put the sensor into the sensor housing.
  • Check if the LED wires and the excitation LED are on the same side.
  • Connect the coax cable to the DAQ and the LED wires with a bit of epoxy glue

  • Also tip each corner of the sensor to the sensor housing.