Skip to content

Multi Measure Munki (mmmunki) is a console utility for working with the X-Rite Colormunki Photo spectrophotometer.

License

Notifications You must be signed in to change notification settings

skorpakov/mmmunki

Folders and files

NameName
Last commit message
Last commit date

Latest commit

 

History

9 Commits
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Repository files navigation

Multi Measure Munki

Multi Measure Munki (mmmunki) is a console utility for working with the X-Rite Colormunki Photo spectrophotometer. It allows you to use this device to measure the color charts used for profiling printing devices and to measure the spectrum and color temperature of light sources.

Table of Contents

Overview

The software that comes with Colormunki Photo has very limited capabilities for generating and measuring color charts: it allows you to use only one-page cards with 50 patches of large size per sheet. Measurement of light sources is not provided at all.

The sensor of the Colormunki Photo spectrophotometer is almost the same as that of the much more expensive i1Pro 2, but its reading frequency is about 10 times less. Therefore, it requires larger patches and overall measurement accuracy is lower. Another drawback of Colormunki Photo is the smaller buffer memory for storing readings. Therefore, if you try to measure a strip with a large number of patches and move the device slowly to increase the number of reads, the buffer may overflow and some data will be lost.

This utility allows you to bypass all these restrictions.

Firstly, it makes it possible to measure any color charts (including multipage) generated in programs that have the ability to save reference tables in text format (Argyll CMS, Profile Maker 5, i1Profiler). The obtained measurement results are also saved in text format. This allows you to use these programs to build profiles of printing devices based on this data.

Secondly, the utility allows using patches of substantially smaller size in color charts without loss of accuracy. The minimum patch size is approximately 11x12 mm, which allows you to place up to 546 patches on an A4 sheet. This is achieved through the ability to measure each strip many times. Then the values that best match certain criteria are selected for each patch.

Thirdly, in order to avoid buffer overflow, the utility allows you to break a strip with patches into fragments and read one fragment per pass. Then the data for each fragment are combined into a strip.

All of this allows to significantly reducing the consumption of expensive paper and ink without loss of measurement accuracy.

In addition, the utility has a special mode that allows you to measure the spectrum and color temperature of light sources with recording data to a file. The obtained data can be used, for example, in an online calculator to calculate the actual color rendition index (CRI) of the light source.

For the convenience of measuring color maps with small patches using a Colormunki Photo spectrophotometer, you can make a simple fixture of cardboard, two rulers and an office file film:

ColorMunki Photo DIY measuring table

Usage

Ambient light measurement mode:  mmmunki -l <measurements file>

Color chart measurement mode:  mmmunki [options] <reference file> <measurements file>
    <reference file> - input reference chart file
    <measurements file> - output measurements file name (without extension)
    options:
      -t - transpose reference data
      -PM5 - save measurements data in ProfileMaker 5 format
      -CB - save measurements data in Epson ColorBase format
      -TI3 - save measurements data in Argyll TI3 format
      -PM5D - save measurements data in ProfileMaker 5 format with density values
      -M<input measurements file1>[,<input measurements file2>[,<input measurements file3> ... ]] - read 
          measurements data from files
      -A<N> - automatically advance to next strip after N measurements (default: 1)
      -D<RGB|CMYK> - save measurements data values in RGB|CMYK format (only supported for 
          ProfileMaker 5 file format)
      -E<76|94|2000> - set delta E calculation method to CIE76|CIE94|CIEDE2000 (default: CIE2000)

Examples

Read reference chart in Argyll CMS format from target.ti2, measure each strip 10 times, save measurements data to result.ti3 file in Argyll CMS format:

mmmunki -TI3 -A10 target.ti2 result

Read reference chart in ProfileMaker 5 format from target_ref.txt, transpose it, measure each strip 10 times, save measurements data to result.txt file in ProfileMaker 5 format:

mmmunki -t -PM5 -A10 target_ref.txt result

Read Epson ColorBase reference chart in ProfileMaker 5 format from SpectroScan.txt, transpose it, measure each strip one time, save measurements data to result.txt file in ColorBase version of ProfileMaker 5 format:

mmmunki -t -CB SpectroScan.txt result

Read reference chart in Argyll CMS format from target.ti2, measure each strip 10 times, save measurements data to result.ti3 file in Argyll CMS format and to result.txt file in ProfileMaker 5 format. This allows you to generate color chart in Argyll CMS, measure it with ColorMunki Photo and create profile in i1Profiler:

mmmunki -TI3 -PM5 -A10 target.ti2 result

If you need to take a break in the middle of measurements, you can save the current results to temporary file, and then load it in the next session:

mmmunki -TI3 -Mtemp.ti3 target.ti2 result

Measure light source spectrum and color temperature and save data to light.txt file.

mmmunki -l light.txt

Now you can analyze it in Spectral calculator. To calculate CRI paste data to text area, open “Additional features and advanced settings” and press “IES-TM-30-15, CRI and CQS” button.

Sample profiling routine using Argyll CMS

Generate preliminary target of one A4 page size (360DPI):

targen -v -w -d2 -G -N 0 -g64 -s16 -B2 -e2 -f546 -c AdobeRGB1998.icc target_pre
printtarg_m -v -t360 -Q8 -ii1 -a 1.08x1.214 -n -P -p 204x286 -L -m0 -z target_pre

Read reference chart in Argyll CMS format from target_pre.ti2, measure each strip one time, save measurements data to target_pre.ti3 file in Argyll CMS format and to target_pre.txt file in ProfileMaker 5 format:

mmmunki -TI3 -PM5 target_pre.ti2 target_pre

Create preliminary profile:

colprof -v -P -D"profile_pre_argyll" -Zr -qh -bh -r1 -al -S AdobeRGB1998.icc -cmt -dpp target_pre

You can also create preliminary profile in i1Profiler using target_pre.txt file.

Check preliminary profiles against measurements:

profcheck -v -w -k target.ti3 "profile_pre_argyll.icm"
profcheck -v -w -k target_pre.ti3 "profile_pre_i1prof.icm"

Generage target of 6 A4 page size using preliminary profile:

targen -v -w -d2 -G -N 0.3 -g256 -s64 -B4 -e4 -f3276 -c "profile_pre_argyll.icm" target
printtarg_m -v -t360 -Q8 -ii1 -a 1.08x1.214 -n -P -p 204x286 -L -m0 -z target

Read reference chart in Argyll CMS format from target.ti2, measure each strip 10 times, save measurements data to target.ti3 file in Argyll CMS format and to target.txt file in ProfileMaker 5 format:

mmmunki -TI3 -PM5 -A10 target.ti2 target

Create profile:

colprof -v -P -D"profile_argyll" -Zr -qu -bu -r1 -al -S AdobeRGB1998.icc -cmt -dpp target

You can also average measurements of preliminary and main color charts:

average -v -m target.ti3 target_pre.ti3 target_plus.ti3

And create profile using averaged data:

colprof -v -P -D"profile_plus_argyll" -Zr -qu -bu -r1 -al -S AdobeRGB1998.icc -cmt -dpp target_plus

And you can create profile in i1Profiler using target.txt file.

Check profiles against measurements:

profcheck -v -w -k target.ti3 "profile_argyll icm"
profcheck -v -w -k target_pre.ti3 "profile_argyll.icm"
profcheck -v -w -k target.ti3 "profile_plus_argyll.icm"
profcheck -v -w -k target_pre.ti3 "profile_plus_argyll.icm"
profcheck -v -w -k target.ti3 "profile_i1prof.icm"	
profcheck -v -w -k target_pre.ti3 "profile_i1prof.icm"

About selecting the best measurement

Two methods have been implemented to select the best measurement. Both are based on calculating the difference in color or delta E.

In the first method, the measurement that is closest (by delta E) to the average value of all measurements is selected. First, an average value is calculated for each color component for all measurements. A delta E is then calculated between these means and each of the measurements. A measurement with a minimum delta E value is selected.

Second method selects measurement with minimum total color difference with all other measurements. Delta E can be considered as the distance between two colors in the color space. For each measurement, the sum of distances to all other measurements is calculated. A measurement is selected with a minimum sum of distances.