Aborting trigger when different modes are selected #103

Closed
prasimix opened this issue Apr 13, 2017 · 3 comments
Closed

Aborting trigger when different modes are selected #103

prasimix opened this issue Apr 13, 2017 · 3 comments
Assignees
Milestone

Comments

@prasimix
Copy link
Member

If for example ch2 trigger is set to FIXed, and ch1 to LIST, switching off ch2 (OUTP OFF) will stop LIST execution on the ch1. There is not need for such behavior despite the fact that single triggering system is used for controlling both channels. Therefore only if channel is in STEP or LIST mode should affect output of other channel when switched off.

@prasimix prasimix added this to the M5 milestone Apr 13, 2017
mvladic added a commit that referenced this issue Apr 18, 2017
@prasimix
Copy link
Member Author

Partially resolved. There is no need that channel's voltage and current output set values stays disabled when channel output is turned off.

mvladic added a commit that referenced this issue Apr 18, 2017
@prasimix
Copy link
Member Author

It's now enabled, but keypad page cannot be entered. Also if encoder confirmation mode is selected it's possible now to "intrude" other channel's (ch1 in our example) output value settings.

mvladic added a commit that referenced this issue Apr 19, 2017
@prasimix
Copy link
Member Author

Fixed