Alarm when expiratory branch is (partially) blocked #45

Open
Benhalor opened this issue Oct 9, 2020 · 2 comments
Open

Alarm when expiratory branch is (partially) blocked #45

Benhalor opened this issue Oct 9, 2020 · 2 comments

Comments

@Benhalor
Copy link
Collaborator

Benhalor commented Oct 9, 2020

The high priority alarm condition indicating that the expiratory branch is partially blocked is not present on the device.
It should be

@dsferruzza
Copy link
Member

We need:

  • alarm code and name
  • priority: high
  • number of respiratory cycle before trigger
  • spec trigger conditions

@Benhalor
Copy link
Collaborator Author

The is alarm is raised when the PEEP is too high