As far as I understood the |
MS3 is, as far as I know, Thermo specific anyway, so it would probably be a good way forward. Matching spectra from different levels is definitely very high priority. |
@jotsetung I have no idea whether The library("mzR")
f <- msdata::proteomics("MS3TMT10_01022016_32917-33481.mzML.gz",
full.names=TRUE)
fh <- openMSfile(f)
header(fh)[c(1, 4, 6),]
# seqNum acquisitionNum msLevel polarity peaksCount totIonCurrent retentionTime
# 1 1 32918 1 1 48304 3005937408 4422.620
# 4 4 32921 2 1 765 1220570 4422.735
# 6 6 32923 3 1 2540 2522363 4423.036
# basePeakMZ basePeakIntensity collisionEnergy ionisationEnergy lowMZ
# 1 696.7169 172620640.0 0 0 376.2217
# 4 549.6967 103048.6 35 0 200.2342
# 6 129.1378 275641.0 65 0 99.0060
# highMZ precursorScanNum precursorMZ precursorCharge precursorIntensity
# 1 1515.1263 0 0.0000 0 0
# 4 1596.3099 32918 673.3353 3 7623700
# 6 505.0462 0 490.2629 0 0
# mergedScan mergedResultScanNum mergedResultStartScanNum
# 1 0 0 0
# 4 0 0 0
# 6 0 0 0
# mergedResultEndScanNum injectionTime
# 1 0 0.0008102598
# 4 0 0.0123253975
# 6 0 0.0520793648
# filterString
# 1 FTMS + p NSI Full ms [380.0000-1500.0000]
# 4 ITMS + c NSI t d Full ms2 [email protected] [180.0000-2000.0000]
# 6 FTMS + p NSI sps d Full ms3 [email protected] [email protected] [100.0000-500.0000]
# spectrumId
# 1 controllerType=0 controllerNumber=1 scan=32918
# 4 controllerType=0 controllerNumber=1 scan=32921
# 6 controllerType=0 controllerNumber=1 scan=32923 As you can see there is always at least a difference of 2 in the scan indices of MS2 and MS3 (but could be even larger). So there is no way to just use the While investigating this problem a little bit more I recognized that the
I don't think we should implement workarounds for But we should provide some high level functions for the generic user to get the parent scans or all children of a current scan, e.g.:
I am not happy with these names because it is hard to understand in which direction they apply the filtering. |
OK, thanks for checking the re filter function names, what about just a single |
|
@sgibb, could you clarify what you mean with
I have a very recent MS3 file, produced on an instrument that has a computer with Xcalibur 3.0.64, and my hopes were still high at that point. I then took that raw file to a computer with the latest proteowizard (version 3.0.11626, freshly downloaded and installed) and converted it. The mzML file then lists Xcalibur 2.1.1565.24 and pwiz 3.0.7364! And, needless to say that the There is an old version of Xcalibur on that computer (version 2.2.44, from June 2011), but as that version doesn't match what I see in the mzML file, I conclude that the local Xcalibur isn't used. For completeness, I also converted the file on another computer that has Xcalibur 3.0.63 (same as on the mass spec computer), pwiz 3.0.999.2 and that mzML file listed Xcalibur 2.1.1565.24 and pwiz 3.0.9987). In a nutshell, proteowizard seems to be way beyond in terms of Xcalibur (or probably MSFileReader) libraries, and isn't internally consistent. God I have proteomics! So, @sgibb, where did you get mzML files that used a recent version of Xcalibur and had these bloody |
Mh, I got that mzML file (with
I have no idea how to tell pwiz to use a different Xcalibur version. Maybe @pavel-shliaha could help us here? |
The computer running the LUMOS machine used to produce the raw file used Xcalibur 4.1.31.9 version. The computer processing the files used: Xcalibur 3.0.63 about the MS3. This can only be performed on ion trap instruments. As far as I know these instruments are currently manufactured by Thermo (LUMOS, FUSION, VELOS, ELITE, Orbitrap XL), Bruker and Shimadzu. Please let me know if you need any more information. Do you perhaps need the raw file in question? |
Thanks @pavel-shliaha. When you say
do you mean that this is what you see in the mzML files? Was Xcalibur installed on that processing computer? If so, what version does it have? Have you done anything specific to get Xcalibur 3.0.63 to be used by pwiz during the conversion? |
yes the "processing computer" is the one used to generate the mzML file and the software installed on it was: Xcalibur 3.0.63 The computer hooked up to the LUMOS instrument that generated the file used 4.1.31.9 version. After the data was acquired, I transferred the data to the processing computer. I have not done anything specific to pwiz, just did normal mzML conversion. Hope this helps. |
Ok, thank you for the clarification. My only explanation is that Xcalibur 4.1.31.9 is needed on the Lumos computer, because otherwise, the setup on our processing computers are pretty much the same. @pavel-shliaha, is that possible based on when you started to get mzML files with complete |
@lgatto if I understand correctly, Sebastian suggests that precursorScanNum is working for MS3 in the file I gave him, but not in other files he tested. And your theory is this is not an export bug of the pwiz or Xcalibur installed on the processing computer, but the problem with the raw files themselves? That could be true, provided we used the very latest Xcalibur software on our LUMOS computers, that was just recently updated (we got the latest version for our UVPD work). It might also be that Thermo fixed the problem earlier when it released SPS on FUSION, since PD (their software) is capable of doing SPS MS3 quantitation, hence it must match MS2 and MS3 spectra. I can look into it, if you haven't yet. I am not sure what instrument was used in the example from msdata |
This is how I explain it as a file acquired today on the Lumos running Xcalibur 3.0.64 and converted using pwiz 3.0.999.2 and having Xcalibur 3.0.64 installed failed to produce a file with a complete
Yes, PD does MS3 quantitation, but they have access to more data in the binary raw files. What really strikes me as utterly weird is that when using the very latest pwiz, the mzML file tells me Xcalibur 2.1.1565.24 was used, which doesn't match any version Xcalibur version (neither on the Lumos computer, neither the processing computer). Any help would be greatly appreciated. |
@sgibb - could you send me that file with the complete |
A couple of things: the software version that is listed "3.0.2022.16" for the file is not the Xcalibur version on either the instrument computer or processing computer. It is the version of the tune, used to generate the data. In fact 3.0.2022.16 is the version of tune installed on our LUMOSes right now. In order to see the tune number from the raw file in Xcalibur. |
Thank you, I think this is what I was missing. On our Lumos computer, is gives 2.1.1565.24. I bet this is why we don't get the |
you have to ask Thermo. However, they might refuse the update, since the new software is beta. We got it since we bought UVPD and 2.X does not support UVPD. |
The wonderful world of fucking mass spectrometry hell. What's UVPD? |
UVPD is ultraviolet photodissociation. Basically you can break molecules by shining light at them of high intensity. This of course requires a laser. We have now had the laser installed inside the instrument see the animation http://players.brightcove.net/665001591001/default_default/index.html?videoId=5449874193001 UVPD (as implemented by Thermo) is mostly beneficial for top-down proteomics, since it takes too long to break peptides (Thermo laser is too weak) |
Sebastian writes that the link to the MS3 file still works. Can you check that. I will also send an MS3 file from the previous version to test https://filesender.deic.dk/filesender/?vid=71c5a478-76c1-4e89-0a7f-000067d3acf9 (should be valid until 2017-12-27) The MS3 file should be called LUM2_01470_KS_L1-5-2_EC17-123-150In2_55oC_NLtrig.raw Please note the MS3 method in this method is NOT a normal TMT MS3 SPS it is a neutral loss trigger for phospho TMT, please see this paper Sensitive and Accurate Quantitation of Phosphopeptides Using TMT Isobaric Labeling Technique PS: I think its the versions of tune again, that are incompatible |
Thanks @pavel-shliaha. Do you possibly have a MS3 and TMT10 or 11 for regular quantitation that was acquired with that recent version of Tune at hand that you could share? |
Thank you for the file @pavel-shliaha! Do you happen to have the corresponsing |
Ping @pavel-shliaha
|
@lgatto. Sorry I was travelling and just arrived yesterday. I searched this data with mascot + percolator in Proteome Discoverer 2.1. It does not return mzIdentML files (I know, since I had trouble submitting to MCP), but it does return a csv file with the quantitation already in place, that can be linked to particular scans. |
Thanks. Could you send me that csv files, please, so that I can extract the information from there. I might still need an |
where do you want the file? I can send it by email. I used human Uniprot canonical and reviewed entries only, I can send it as well |
Both by email is fine, thanks. |
Keeping open to add an example and section in the vignette using new data to be added to |
Last week @pavel-shliaha ask me to check whether a new neutral loss method for phosphopeptides is working as expected. Therefore it was important to find out which MS3 spectrum was triggered by which MS2. I did that by matching the precursorMz used for the MS2 scan from the filterString (the MS3 scan filterString contains the precursor for MS2 and MS3; that was one of the reasons for implementing it into
mzR
):I did the matching as follows:
See the whole example at https://rawgit.com/sgibb/talk-odense-20171127/master/01-neutral-loss.html#compare-nl-trigger (unfortunately it doesn't contain the dataset because it is 2.2 Gb but @pavel-shliaha promise me that he would share this dataset if needed).
While this procedure isn't very complicated we should provide a function or something else to allow an easier reference for each MSn to its corresponding MS(n-1).
I am not sure that the filterString approach will always work. Maybe we have to go through all precursorMz or some other information. Also I am not sure whether it would be enough to add another column to the
featureData
slot (e.g.parentMsScanId
) or we need to add it to theSpectrum
class.Do you have any better ideas? Do you thing we need such a kind of reference?
The text was updated successfully, but these errors were encountered: